To prevent an electrical shock hazard to the operator and/or damage to the instruments, 

read this instruction manual carefully before using the instrument. 

WARNINGS

 with the 

symbol       on the instrument and this instruction manual are highly important.

Important Symbols

The symbol in this manual advises the user of an electrical shock 

hazard that could cause injury or material damages.

  

1. DISPLAY

LCD

    

a. Numerical Display

 

4 digit 6000 count, with Bar Graph

    

b. Units and Symbols

 

 2. OPERATING PRINCIPLE

 Sigma-delta Conversion

  3. MEASURING PRINCIPLE

 

Average Rectification

 

 4. SAMPLING RATE

 

5 times per second (Bar Graph

40 times per second max.)

 

 5. RANGE SELECTION

Auto / Manual (Frequency, Injection Pulse width and

Capacitance measurement are Auto only)

 

 6. BATTERY WARNING

:      

indication at approx. 2.3V or less

  

7. DISPLAY HOLD

Hold indicating values by pressing HOLD Key

  8. AUTO POWER OFF

Power turns off automatically after approx. 34 minutes  

  9. OVERLOAD PROTECTION

10. OPERATING TEMPERATURE & HUMIDITY

Maximum relative humidity 80% for temperature up to 31

 decreasing linearly to 50% 

relative humidity at 40

 

11. STORAGE TEMPERATURE & HUMIDITY

       -20 to 60

less than

 

80%RH in non-condensing.

12. TEMPERATURE COEFFICIENT

Accuracy at 23

±

5

×

0.15/

13. SAFETY LEVEL

CE marking approved.

14. POWER SUPPLY

1.5V R03(AAA) Battery 

×

2

15. FUSE

Fast-acting 11A/1000V

 (

φ

10

×

38mm

) ×

1

Fast-acting 0.4A/1000V

 (

φ

6

×

32mm

) ×

1

16. POWER CONSUMPTION

approx.

 

4.3mA

(in Auto Power Off 

approx. 10

μ

A, in backlighting

approx. 27mA)

17. CONTINUOUS OPERATING TIME

approx. 180 hours (manganese cell),

approx. 360 hours (alkaline cell)

18. DIMENSIONS & WEIGHT

161(H)

×

80(W)

×

50(D)mm, approx. 340g

19. ACCESSORIES

20. OPTIONAL ACCESSORIES

a) 

μ

A & mA

0.4A/1000V fast-acting fuse

b)  A

11A/1000V fast-acting fuse

c)  V

1050Vrms or 1450Vpeak

d)  mV, 

Ω

 and others

1000Vrms

Thank you for purchasing 

"KT-2022 AUTOMOTIVE DIGITAL MULTIMETER"

. To obtain the maximum 

performance of this instrument, read this Instruction Manual carefully, and take safe measurement.

2. SPECIFICATIONS

Do not measure High Power Line (High Energy Circuits) that might exceed 6kVA. High 

Power Line is very dangerous and sometimes includes High Surge Voltage that could 

cause explosive short in the instrument and could result in serious injury to the operator. For 

dangerous voltage measurement, always keep the instrument away from your body without 

holding it in your hands.  Do not touch the instrument, test leads, and any part of the circuit.

WARNING

FOR SAFETY MEASUREMENTS!!

The symbol listed in IEC 61010-1 and ISO 3864 means "Caution (refer to 

instruction manual)".

WARNING

CAUTION

The symbol in this manual advises the user of an electrical shock 

hazard that could result in serious injury or even death.

INTRODUCTION

Confirm if the following items are contained in the package in good condition.

If there are any damage or missing items, ask your local dealer for replacement.

1. Digital Multimeter (with holster)

2. Test Lead (100-66)

3. RPM Sensor (653 / for Direct Ignition Engine)

4. RPM Sensor (654 / for High Tension Code)

5. Temperature Probe (818-02)

6. Alligator Clip (943)

7. Carrying Case (1024)

8. Spare Fuse (F38, 11A/1000V and F39, 0.4A/1000V)

9. Instruction Manual

1. UNPACKING AND INSPECTIONS

1 pce.

1 set

1 pce.

1 pce.

1 pce.

1 pce.

1 pce.

1 pce. each

1 pce.

AUTO,      ,     ,     ,     , RPM     , RPM     ,     ,     , 

Ω

, k

Ω

, M

Ω

,

Hz, kHz, ms, V, 

μ

F, 

μ

A, mA, A, 

, Trig and decimal point

2-1. GENERAL SPECIFICATIONS

2-2. MEASUREMENT SPECIFICATIONS

 

(23

±

5

, <80

RH in non-condensing)

1. DC Voltage  (     mV / V )

6.000 V

60.00 V

600.0 V

1000 V

±

0.4

rdg

±

3dgt 

±

0.7

rdg

±

3dgt 

Range

60.00 mV

600.0 mV

10M

Ω

,

50pF

1000Vrms

Accuracy

±

0.4

rdg

±

3dgt 

±

0.3

rdg

±

3dgt 

Resolution

0.01 mV

0.1 mV

Input

Impedance

Overload

Protection

10M

Ω

,

50pF

1050Vrms

1450Vpeak

1 mV

10 mV

100 mV

1 V

2. AC Voltage  ( 

mV / V )    50Hz to 500Hz

Average Rectification

3. DC Current  (    

μ

A / mA / A )

10A Continuous measurement is possible.

10A to 20A is measurable within 30 seconds but 5 minutes cooling down interval is necessary.

4. AC Current  ( 

μ

A / mA / A )   50Hz to 500Hz

10A Continuous measurement is possible.

10A to 20A is measurable within 30 seconds but 5 minutes cooling down interval is necessary.

Average Rectification

5. Resistance  ( 

Ω

 )

Range

600.0 

Ω

6.000 k

Ω

60.00 k

Ω

600.0 k

Ω

6.000 M

Ω

60.00 M

Ω

0.45V DC

Accuracy

±

0.5

rdg

±

6dgt 

±

0.5

rdg

±

3dgt 

±

0.8

rdg

±

4dgt 

±

1.0

rdg

±

5dgt 

±

1.5

rdg

±

5dgt 

Resolution

0.1 

Ω

Ω

10 

Ω

100 

Ω

1 k

Ω

10 k

Ω

Open Circuit Voltage

6. Frequency  ( Hz )

Function

6 V

60 V

600 V

1000 V

±

0.1

rdg

±

3dgt

Range (Auto)

10 Hz to 10 kHz
10 Hz to 50 kHz

45 Hz to 1 kHz

Sensitivity

0.5 V rms

5 V rms

50 V rms

500 V rms

Accuracy

NOTE :

 Measurement cannot be made depending on the vehicles.

RPM4

RPM2

RPM2-M

Range

240 to 20000RPM

120 to 10000RPM

60 to 5000RPM

7. IP-RPM ( Inductive Pickup type )

±

0.2

rdg

±

20dgt

Accuracy

NOTE :

 Measurement cannot be made depending on the vehicles.

RPM4

RPM2

RPM2-M

Range

60 to 20000RPM

30 to 10000RPM

15 to 5000RPM

8. IG-RPM ( Contact Signal type )

±

0.2

rdg

±

20dgt

Accuracy

Selectable Number of Cylinders

1,2,3,4,5,6,8,10,12

(Select by 

CYLINDER Key

)

      WARNING 1. Checks of Body and Test Lead

      WARNING 3. Warning for High Voltage Measurements

      WARNING 4. Dangerous Voltage Measurement Procedure

      WARNING 2. High Power Line Measurements is Prohibited  

Before measurement, confirm the body of this instrument and handle insulators of the Test 

Lead have no cracks or any other damages. Dust, grease and moisture must be removed.

Do not hold multimeter in your hands. Keep safety distance from power source or circuit 

to be measured not to touch the dangerous voltage.

Attach black and red alligator clips to test lead pins.

Turn off the power of the circuit to be measured when connecting test leads.

After measurement, before detaching alligator clips (test leads), turn the circuit power off 

and discharge the all capacitors.



For dangerous voltage measurement, strictly observe the warnings below.

3-1. WARNINGS

Correct knowledge of electric measurements is essential to avoid unexpected danger such 

as operator's injury or damage to the instrument. Read carefully and observe the following 

precautions for safety measurements.

3. SAFETY PRECAUTIONS

Even for Low Energy Circuits of electric/electronic appliances, such as heating elements, 

small motors, line cords and plugs, High Voltage Measurements are very dangerous. Do not 

touch any part of the circuit. 

Do not measure High Power Line (High Energy Circuits) such as Distribution Transformers, 

Bus Bars and Large Motors. High Power Line sometimes includes High Surge Voltage that 

could cause explosive short in the instrument and could result in shock hazard. Generally, 

shock hazard could occur when the current between the circuit, that involves more than 33V 

rms or 46.7V DC or peak, and ground goes up to 0.5mA or more.

      WARNING 1. Correct Selection of Function Switch

      WARNING 2. Maximum Input Observance

      WARNING 3. Test Lead Detachment

Do not measure anything that might exceed the specified maximum input values.

Always confirm that 

FUNCTION

 Switch is set to the correct position. Do not measure voltage 

except at Voltage measurement function.

Detach test leads from measuring circuit before changing measurement functions or 

removing rear cover for battery or fuse replacement.

In case of live-line measurement, strictly observe the warnings below 

Do not hold multimeter in your hands. Keep safety distance from power source or circuit 

to be measured not to touch the dangerous voltage.

Black test lead : Attach black alligator clip and connect to 

 (earth) side of the circuit.

Red test lead : Connect to 

 (positive) side of the circuit.


3-3. GENERAL WARNINGS AND CAUTIONS

Children and the persons who do not have enough knowledge about 

electric measurements must not use this instrument.

Do not measure the electricity naked or barefooted to protect yourself 

from electrical shock hazard.

Be careful not to get hurt with the sharp test lead pins.

Do not polish the case or attempt to clean it with any cleaning fluid like 

gasoline or benzine. If necessary, use silicon oil or antistatic fluid.

Away the instrument from hot and humid conditions like in the car. Do not 

apply hard mechanical shock or vibration.

Remove the battery when the instrument is out of use for a long time. The 

exhausted battery might leak electrolyte and corrode the inside.

WARNING 1.

 

WARNING 2.

 

WARNING 3. 

CAUTION 1.

CAUTION 2. 

CAUTION 3.

CAUTION  4.         

 

Power turns off automatically after approx. 34 minutes of last operation.

To cancel it : 

Turn the power on holding down 

Select 

Key. Auto power off is disabled.

NOTE :

 Small current consumption remains even in the auto power off condition.

Be sure to set 

FUNCTION

 Switch to 

"OFF" 

after finishing the measurement.

Caution (refer to instruction manual)

Alternating Current

Direct Current

Earth (Ground)

Fuse

Double Insulation

DC / AC

Do not apply around, or remove from HAZARDOUS LIVE conductors.

LCD displays "

OL

" when measurement value exceeds the maximum value of each 

measurement range.

The following symbol marks shown on the instrument and instruction manual are listed in 

IEC 61010-1 and ISO 3864.

0.4A/1000V and 11A/1000V fuses are installed to protect current measurement functions. If 

the fuse is blown out, replace it referring to "

7-1. BATTERY AND FUSE REPLACEMENT

".

4. NAME ILLUSTRATION

fig-1

LCD

FUNCTION Switch

Input

Terminals

Hold Key

±

TRIGGER Key

(

Key

CYLINDER Key

(

Select Key

     Key

(

Range Key

Level Key

(

fig-2

4-1. LCD

: 

Auto-ranging

: 

Low battery warning

: 

Dwell angle measurement

: 

Continuity test

: 

Lights up

      in display hold function

: 

RPM measurement

: 

Direct current (DC)

RPM

: 

Alternating current (AC)

: 

Resistance measurement

: 

Frequency measurement

: 

Injection pulse width measurement

: 

Voltage measurement

: 

Capacitance measurement

: 

Current measurement

: 

Duty cycle

Ω

, k

Ω

, M

Ω

Hz, kHz

ms

mV, V

μ

F

μ

A, mA, A

4-11. Input Terminals

4-10. Level Key

Use this key to select the sensitivity level in IP-RPM, IG-RPM, Dwell Angle and Injection 

Pulse Width measurements.

Connect black test lead to COM terminal and red test lead to RPM/Temp+/

Ω

/V/     Terminal 

for all measurements except 

μ

A, mA, A measurements.

NOTE

Connect red test lead to mA/

μ

A terminal for mA/

μ

A measurement and connect it to

A terminal for 6A (10A for 30sec) measurement.

5. MEASUREMENT PROCEDURES

5-1. PREPARATION FOR USE

Read INSTRUCTION MANUAL carefully to understand the specification and functions 

correctly. "

3. SAFETY PRECAUTIONS

" is very important for safety measurement.

1. INSTRUCTION MANUAL

2. BATTERY

Two 1.5V R03 batteries are installed in this instrument. When "      " lights up on LCD, replace 

them into the new ones referring to "

7-1. BATTERY AND FUSE REPLACEMENT

".

3. FUSE

4. OVERLOAD INDICATION

5. AUTO POWER OFF

6. SYMBOL MARK

Specified ranges depend on engine rpm and number of cylinders (cyl).

Range

0.0

°

 to 360.0

°

0.0

 to 100.0

Accuracy

±

1.2

°

/krpm

±

1dgt 

±

0.04

/krpm/cyl

±

2dgt 

9. Dwell Angle (      ), Duty Cycle ( 

 )

Selectable Number of Cylinders

1,2,3,4,5,6,8,10,12

(Select by 

CYLINDER Key

)

Specified ranges depend on trigger slopes, engine rpm and number of cylinders (cyl).

Accuracy

±

0.05ms

±

1dgt

±

0.04%/krpm

±

2dgt

±

0.05ms

±

1dgt

±

0.04%/krpm/cyl

±

2dgt

Range

0.05ms to 250.0ms

0.0% to 100.0%

0.05ms to 250.0ms

0.0% to 100.0%

10. Injection Pulse Width ( ms ),  Duty Cycle ( 

 

)

Multi Point Injection

Single Point Injection

WARNING

Do not measure any voltage that might exceed maximum input value.

(V

1000V AC/DC, mV

600mV AC/DC). 

Read "

3. SAFETY PRECAUTIONS

" carefully to avoid electric shock hazard and 

serious damage to the instrument.

Do not touch any part of power line or the circuit to be measured.

1

2

3

4

5

6

Connect the instrument 

IN 

PARALLEL

 to the circuit.

Use alligator clips for dangerous 

voltage measurement.

NOTE

Display shows "   V

→〜

V

Hz " 

each time the 

Select

 Key is pressed.

Connect black test lead to 

 (earth) side 

of the circuit being measured and connect 

red test lead to 

 (positive) side.

NOTE :

 

NOTE :

Read the measurement value on LCD.

After finishing the measurement, set 

FUNCTION

 Switch to 

"OFF"

.

5-2. VOLTAGE AND FREQUENCY MEASUREMENT (    V / 

V / Hz ) (    mV / 

mV)

fig-3

Battery voltage

measurement

Black(

)

Red(

)

Battery

(    V / 

V / Hz )

Insert black test lead to 

COM

 terminal, and insert red test lead to 

RPM/Temp

/

Ω

/V/

terminal.

Set 

FUNCTION

 Switch to "       ".

Press 

Select

 Key to select DC(    ) or AC(    ).

V

Hz

100-66 Test Lead, 653 RPM Sensor (for Direct Ignition), 654 RPM Sensor (for High 

Tension Code), 818-02 Temperature Probe, 943 Alligator Clip, 1024 Carrying Case, 

Holster, 1.5V R03(AAA) Battery (Installed), F38 Spare Fuse (11A/1000V),  F39 Spare 

Fuse (0.4A/1000V), Instruction Manual

11. Diode Test  (        )

Range

1.000 V

Test Current

0.50mA typ.

Open Circuit Voltage

1.6V DC typ.

Accuracy

±

1.0

rdg

±

3dgt 

Range

6.000

μ

F

60.00

μ

F

600.0

μ

F

2000

μ

F

Accuracy

±

2.0

rdg

±

5dgt 

±

3.5

rdg

±

5dgt 

±

4.0

rdg

±

5dgt 

14. Capacitance  (        )

660 AC/DC Clamp Adapter, 100-41 Test Lead Kit, 100-62 Test Lead Set, 944 Test Pin, 

946 Battery Clip, 793 Coil Contact Pin

Threshold Level

between 10

Ω

 and 200

Ω

Response Time

32ms

13. Continuity  (       )

Turns the power on and selects measurement functions. After measurement, turn it to 

"OFF"

.

4-3. Select Key

Use this Key to select sub-measurement functions.

4-2. FUNCTION Switch

Press this key for 1 second or more to turn on the LCD backlight. Press for 1 second or more 

again, backlight is turned off.

4-4. 

  

Key

4-6. CYLINDER Key

4-8. 

±

TRIGGER Key

Change 

/

 Trigger slopes in Duty Cycle and Injection Pulse Width measurements. (Refer to 6-2).

4-5. 

   

Key

4-7. Hold Key

Display Hold

Use this key to select the engine strokes in IP-RPM and IG-RPM measurements: 

Four-stroke/Two-stroke/Two-stroke, 2-cylinder simultaneous ignition.

Use this Key to set number of cylinders in Dwell Angle and IG-RPM measurements.

Holds displayed value on LCD by pressing this Key. ("

  

" lights up).

To release it

Press 

Hold 

Key again.

FREQUENCY MEASUREMENT  ( Hz )

(1)

(2)

(3)

Press 

Select

 Key during voltage measurement. "Hz" lights up on LCD and Frequency 

of the measuring voltage is displayed.

Press 

Select

 Key again to return to voltage measurement.

In Frequency measurement, input sensitivity can be changed in the following procedures.

a.

b.

c.

Press 

Level

 Key during frequency measurement. Input sensitivity level is changed 

as Level 1 (6V)

2 (60V)

3 (600V)

4 (1000V). (default setting is Level 1 (6V)).

Sensitivity level is shown with the number of "    " under the bar graph.

Change of input sensitivity is effective in the following cases.

When measured value is unstable, reduce the current input sensitivity level.

When measured value is "0", increase the current input sensitivity level.

(    mV / 

mV )

1

2

3

4

5

6

Insert black test lead to 

COM

 terminal, and insert red test lead to 

RPM/Temp

/

Ω

/V/

terminal.

Set 

FUNCTION

 Switch to "           ".

Press 

Select

 Key to select DC (     ) or AC (     )

NOTE

Display shows "     mV

→〜

mV

→°

C

→°

F " each time the 

Select

 Key is pressed.

Connect black test lead to 

 (earth) side of the circuit being measured and connect red 

test lead to 

 (positive) side.

NOTE : 

Connect the instrument 

IN PARALLEL

 to the circuit.

NOTE : 

Use alligator clips for dangerous voltage measurement.

Read the measurement value on LCD.

After finishing the measurement, set 

FUNCTION

 Switch to 

"OFF"

.

Do not measure voltage in "                           " position. This will cause electrical shock 

hazard to the operator and/or damage to the instrument.

Be careful not to drop off the instrument and test leads into the engine room.

Ignition system generates high voltage that could cause an electrical shock hazard. 

Stop the engine when connecting or removing the test leads.

Read "

3. SAFETY PRECAUTIONS

" carefully before measurement.


WARNING

5-3. RPM MEASUREMENT  (IG-RPM)  

measuring RPM with test lead

-ms

IG-RPM

Stop the engine.

Insert black test lead to 

COM

 terminal, and insert red test lead to 

RPM/Temp

/

Ω

/V/

terminal.

Set 

FUNCTION

 Switch to "                       ", and press 

Select

 Key until "RPM" appears 

on LCD.

-ms

IG-RPM

1

2

3

4-9. Range Key

Use this Key to change Auto or Manual-ranging (Refer to 6-1).

3-2. PREVENTION OF FAILURE

Maximum Input
600mV AC/DC

1000V AC/DC

6.000 V

60.00 V

600.0 V

1000 V

±

2.0

rdg

±

5dgt

±

2.2

rdg

±

5dgt

Range

60.00 mV

600.0 mV

10M

Ω

,

50pF

1000Vrms

Resolution

0.01 mV

0.1 mV

Input

Impedance

Overload

Protection

10M

Ω

,

50pF

1050Vrms

1450Vpeak

Maximum Input
600mV AC/DC

1000V AC/DC

Accuracy

±

2.0

rdg

±

5dgt 

1 mV

10 mV

100 mV

1 V

60.00 mA

600.0 mA

6.000 mA

10.00 mA

Range

600.0 

μ

A

6000 

μ

A

0.25mV

μ

A

0.4A

1000V

Fuse

Resolution

0.1 

μ

A

μ

A

Burden Voltage

Overload

Protection

Maximum Input

6000

μ

A AC/DC

0.01 mA

0.1 mA

1 mA

10 mA

Accuracy

±

0.7

rdg

±

3dgt 

±

0.5

rdg

±

3dgt 

±

0.7

rdg

±

3dgt 

±

0.5

rdg

±

3dgt 

±

0.7

rdg

±

3dgt 

±

0.5

rdg

±

3dgt 

2.5mV

mA

0.4A

1000V

Fuse

600mA AC/DC

0.03V

A

11A

1000V

Fuse

10A AC/DC

60.00 mA

600.0 mA

6.000 mA

10.00 mA

Range

600.0 

μ

A

6000 

μ

A

0.25mV

μ

A

0.4A

1000V

Fuse

Resolution

0.1 

μ

A

μ

A

Burden Voltage

Overload

Protection

Maximum Input

6000

μ

A AC/DC

0.01 mA

0.1 mA

1 mA

10 mA

Accuracy

±

2.2

rdg

±

5dgt 

±

2.0

rdg

±

5dgt 

±

2.2

rdg

±

5dgt 

±

2.0

rdg

±

5dgt 

±

2.2

rdg

±

5dgt 

±

1.2

rdg

±

5dgt 

2.5mV

mA

0.4A

1000V

Fuse

600mA AC/DC

0.03V

A

11A

1000V

Fuse

10A AC/DC

1000Vrms

Overload Protection

6553 count max.

Display

1000Vrms

Overload

Protection

1000Vrms

Overload Protection

Resolution

0.001

μ

F

0.01

μ

F

0.1

μ

F

F

1000Vrms

Overload Protection

KT-2022

AUTOMOTIVE

DIGITAL MULTIMETER

INSTRUCTION MANUAL

LVD

CAT

1000V, CAT

600V, CAT

300V AC and DC

EMC

EN61326-1 (2013)

When using Clamp Adapter, read its instruction manual carefully.

K-type thermocouple / sensor accuracy is not included.

Range

-

50

°

C to 1000

°

C

-

58

°

F to 1832

°

F

Accuracy

±

0.5

rdg

±

3dgt 

±

0.5

rdg

±

6dgt 

12. Temperature  ( 

°

C

 

 

°

F )

Resolution

1

°

C

1

°

F

1000Vrms

Overload Protection