U3022AM06
Cal Kit Operational Check
Cal Set Verification
1.
Select
Trace/Chan > Trace > Delete Trace.
There should be no traces on the
display.
2.
To launch the Cal Set Viewer toolbar. Select
Response > Cal > Manage Cals > Cal
Set Viewer
.
Figure
60
Calibration, Cal Set Viewer
3.
From the Cal Sets drop-down menu, select
999.1
and select
Enable
. Select the
Reflection Tracking(x,x)
,
where x,x is the port being tested. Ensure that the
Enable
and
Error Terms
check boxes are selected.
4.
Compare the Reflection Tracking (1,1) trace to the appropriate limits in
. The trace should be above the limit.
5.
Repeat
step 3
and
step 4
for Cal Sets "999.1" thru "999.8" (8-Port).
74
Keysight U3022AM06 User's and Service Guide