![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 66](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380066.webp)
U3022AM06
Control Lines
Internal Voltage Supply Configuration
The output voltage of pin 12 can be adjusted from +2 to +5 V. Perform the following
procedure to set the voltage:
1. Turn on the Test Set.
2. Measure the voltage between pin 12 and 15 using a multimeter.
3. Rotate the voltage adjustment trimmer on the rear panel until the multimeter
indicates the selected voltage
Figure 5
6
illustrates an example of the connection between the DUT and the Test Set
using
the internal DC power supply. Connect pin 12 to pin 13 and pin 14 to pin 15 to
provide the ground path. Connect the control lines to the external DUT.
You may only connect pin 12–13, and pin 14–15, damage may result if any other
paths are short-circuited.
Figure 5
6
Internal DC Power Configuration (rear panel view)
9
8
1
15
14
13
12
TO DUT (Only required Lines)
GND
12
13
14
15
1
2
8
2 to 5 Vdc
TEST SET
Line 1
Line 2
Line 8
GND
(Rear Panel View)