U3022AM06
General Specifications
Environmental Requirements
Refer to your Analyzer’s standard documentation for environmental requirements.
The Test Set complies with all applicable safety and regulatory requirements for the
intended location of use.
•
Pressure Altitude (Operation)
3,000 meters (~10,000 feet)
•
The instrument can safely operate in a relative humidity of 80% for temperatures
to 31 degrees C, decreasing linearly to 50% relative humidity at 40 degrees C.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature
within the appropriate range.
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If
the total power dissipated in the cabinet is greater than 800 watts, forced convection
must be used.
Table 4
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient
temperature of the Analyzer and Test Set to be maintained within ± 1 °C of the
ambient temperature at calibration.
Dimensions and Space Requirements
Standard installation of the Test Set and Analyzer includes configuration and
installation on a customer provided lab bench, or table top of adequate size and
strength. For weight, dimensions and space requirements, refer to the network
analyzer documentation that you are using to configure the Test Set.
Instrument Dimensions
Model
Weight
Height
Width
Depth
U3022AM06
11.4 kg (25 lb)
19.1 cm
(7.5 in)
42.5 cm
(16.7 in)
43.2 cm
(17 in)