![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 59](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380059.webp)
U3022AM06
Controlling the Test Set
Figure 52
Method 2 - Using Test Set Address and Data Values
This first command creates a new S10_10 measurement on channel 1 of the PNA-X, named
“ch1_S10_10,” and configures the internal Test Set RF switches.
CALC1:PAR:EXT 'ch1_S10_10', 'S10_10'
This second command feeds the newly created measurement named “ch1_S10_10” to trace
2 on the PNA-X so that it will be displayed on the PNA-X screen.
DISP:WIND1:TRAC2:FEED 'ch1_S10_10'
NOTE: Here are syntax format examples for single digit S-parameters:
‘ch1_S99’ ‘ch1_S22’ ‘ch1_S9_10’ ‘ch1_S10_9’
Method
2
- Using
the Test Set Address and Data
Values
This method is available while the system is in
Standalone
mode only. The
address and data
values for the Test Set can be found in the
"Address and Data Values" section on page 60
The Test Set internal switch settings are programmed into the firmware of the Test Set and
can be controlled with the specific Address and Data values. Each Address and Data value
pair sets the Source or Receiver switch paths. For an S-parameter measurement, two pairs of
Address and Data values will be needed.
Address and Data values are separated by a comma. Commands should be separated by a
new line, or carriage return.
For example:
CONT:EXT:TEST:DATA <address>,<data>
CONT:EXT:TEST:DATA 0,0
Example: CONT:EXT:TEST:DATA 0,0
Keysight U3022AM06 User's and Service Guide
59