Line Module
Available Fuses
U3022AM06
Front and Rear Panel Features
This assembly houses the line cord connection, line fuse, and line voltage selector.
Remove the line module cover to replace or change the fuse. Line voltage selection is
automatic and no setting is required.
Install the instrument so that the detachable power cord is readily identifiable and is
easily reached by the operator. The detachable power cord is the instrument
disconnecting device. It disconnects the mains circuits from the mains supply before
other parts of the instrument. The front panel switch is only a standby switch and is
not a LINE switch. Alternatively, an externally installed switch or circuit breaker
(which is readily identifiable and is easily reached by the operator) may be used as a
disconnecting device.
Fuse (F 5 A/250V, 2110-0709) UL listed and CSA certified
For continued protection against fire hazard replace line fuse only with same type
and rating. The use of other fuses or material is prohibited.
Figure 8
Line Module and Fuse
Verify that the premise electrical voltage supply is within the range specified on the
instrument.