![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 17](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380017.webp)
U3022AM06
General Specifications
Frequency Range and Power Levels
When the Test Set is configured with your network analyzer, the system frequency
range of operation will be determined by the frequency range limits of this
combination.
It is recommend that you do not operate components near damage levels
(+30 dBm, 7 Vdc). The power levels must be 3 dB below maximum level to ensure no
damage.
Refer to your Analyzer specifications to determine the maximum input power levels
for the Analyzer access and test ports, or to optimize the power levels in the
receivers.
Damage and maximum levels are not necessarily the optimum level.
Table 6
Recommended Maximum Power Levels and Vrms
a
a.
Formula for 50 Ω system. P(dBm) = 10 Log
10
(V
rms
2
/0.05)
U3022AM06
Test Port RF Power Levels
PORT 3-8
+27 dBm 5 Vdc
U3022AM06
Access Ports:
SOURCE OUT
+20 dBm 2.2 Vdc
CPLR ARM
+20 dBm 2.2 Vdc
CPLR THRU
+20 dBm 2.2 Vdc
RCVR OUT
+20 dBm 2.2 Vdc
Table 5
Usable Frequency Range
Instrument
N5230A/C (PNA-L)
300 kHz to 20 GHz
Frequency Range
N5242A/B (PNA-X)
10 MHz to 26.5 GHz
E8362B/C
10 MHz to 20 GHz
Keysight U3022AM06 User's and Service Guide
17