![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 45](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380045.webp)
U3022AM06
Controlling the Test Set
Multiport Mode (Option 551)
Multiport mode selects the Test Set file that will enable the network analyzer to
control the Test Set. Multiport mode allows you to complete a N-Port calibration
using the Cal Wizard application in the network analyzer. Refer to the Help system for
more information.
Ensure the Test Set file needed for your analyzer are installed into network analyzer
file directory: C:\Program Files\
Keysight
\Network Analyzer\testsets
“Network Analyzer Requirements” on page
How to Access Multiport Mode
1.
The Option 551 must be installed for multiport capability. To access the multiport
application select Utility > System > Configure > Multiport Capability
- OR -
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...)
.
2.
Select U3022AM06 (8 or 10-Port System) from the drop-down menu and select
Restart as a multiport PNA with this testset > OK. The Analyzer will restart the
network application with the Test Set interface features.
Figure 35
If the Test Set is not available in the drop-down list, it will be necessary for you to
copy the required Test Set file to the Analyzer’s hard drive. The current version of the
Test Set files are available on the web at
http://na.support.keysight.com/multiport
.
Copy the appropriate file to c:\Program Files\
Keysight
\Network Analyzer\testsets
directory.
Selecting Multiport Mode
Keysight U3022AM06 User's and Service Guide
45