U3022AM06
Cal Kit Operational Check
Verifying Cal Kit Operational Check Failure
If your test results fail the Cal Kit Operational Check limits, see
and
and verify the following:
1.
Ensure that the Test Set is turned On and connected to the analyzer.
2.
Check all appropriate analyzer and Test Set connectors for damage, cleanliness,
and proper torque.
3.
Repeat the relevant 1-Port calibrations using another ECal or mechanical
standard.
4.
Verify that the analyzer is operating properly and meeting its published
specifications. If the analyzer is connected to the Test Set it will fail the
Operational Check. Refer to
“Network Analyzer Requirements” on page
ECal Confidence Check
The following procedure is a method to check the accuracy of a 1-Port calibration
performed with mechanical standards or an ECal module. In order to perform this
check, the test port of the ECal module must connect directly to the test port being
verified (without adapters).
1.
Perform a 1-Port calibration on the Test Set port being tested.
2.
Connect the ECal Module to the port being tested. Terminate any remaining ports
on
the ECal Module.
3.
Select the Cal Set to be tested,
[CAL] > Cal Set > Cal File > Apply Cal > Close
-OR-
([Cal] Cal Set & Cal Kit > Cal Set... > Apply Cal)
.
If
the Choose Stimulus Settings
prompt appears, select
Change the active channel's stimulus > OK > Close
.
4.
Select
[Response] > Cal > More > ECal > ECal Confidence Check
-OR-
(Response > Cal > ECal Confidence Check...)
.
5.
ECal Confidence Check dialog box: Click
Change Measurement
and select the test
port
S-parameter > Apply > OK
. Click
Read Module Data
.
6.
Select
ECal Module
dialog box: Select the ECal Module you are using, and select
the
ECal Module Memory and Factor ID > OK
.
Figure 6
4
ECal Confidence Check
Keysight U3022AM06 User's and Service Guide
77