Figure 37
U3022AM06
Controlling the Test Set
Trace Measure S-Parameter
S-Parameter selection can be accomplished using Response > Measure. Use the
drop-down menus to select 1 of 100 S-Parameters for the 10-Port system,. The first
number in the Sxx selection is the Receiver Port and the second number will be the
Source Port. Any port can be selected to be the Receiver, Source or both, as in S11.
The front panel R LED indicates the port is the Receiver and the S LED indicates the
port is the Source.
The 'B' Model analyzer provides an S-parameter selection using the keypad. For
example, to select S25:
MEAS [MEAS] > Enter S-Parameters > [2][.][5] [Enter]
.
S-Parameter Measurement
Figure 38
New Trace Measure S-Parameter
S-Parameter Tab: Multiple S-Parameters can be made from the New Trace menu. In
the drop-down menu select Trace/Chan > Trace > New Trace.
- OR-
(Instrument > Trace > New Trace)
.
The dialog box allows the selection of any of the
.
New S-Parameter Measurement
Keysight U3022AM06 User's and Service Guide
47