![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 50](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380050.webp)
U3022AM06
Controlling the Test Set
RF Path Configuration with Option 029
If the analyzer has Option 029 or H29 (noise figure capability) verify that the path
configuration is set to Default. If it is not in Default configuration, Port 5 will not
operate correctly.
1.
Select
Trace/Chan > Channel > Hardware Setup > Path Config...
-
OR-
(Instrument > Setup > Internal Hardware > RF Path Config...)
and in the drop-down
menu select Default > OK. Save this configuration as a User Preset by
selecting
Save > User Preset > Save current state as User Preset
. Do not use the
factory
Preset
(User Preset Off), the analyzer will return to Option 029 path
configuration.
2.
Verify that the Port 1 Noise Tuner Switch is set to external. Select
Utilities System>
Configure > Preferences
. If not, select
Meas: Port 1 Noise Tuner Switch is set to
external
.
Figure 43
RF Path Configuration
50
Keysight U3022AM06 User's and Service Guide