![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 72](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380072.webp)
U3022AM06
Cal Kit Operational Check
Operational Check Procedure
The sequence of this procedure is very important and must be followed or the
performance accuracy and results may vary from the reference plots provided. Ensure
that the Test Set is not connected to the analyzer if you are performing an Operator’s
Check. The analyzer will indicate false failures if the Test Set is connected.
On a 10-Port configuration, Ports 9 and 10 are not connected to the Test Set;
therefore, only Ports 1 thru 8 are tested.
If you suspect that your 8 or 10-Port configuration is not operating properly, ensure
that all front RF jumper interconnect cables are correctly attached.
Preparing the Network Analyzer
1.
Connect the Test Set to the analyzer using the interconnect cables as shown in
Table
2
“Interface Kit Options” on page
2.
Turn
on
the Test Set.
3.
Select
Response > Cal > Manage Cals > Cal Set
.
Delete or rename any Cal Sets titled
“999.1” thru “999.8” (8-Port), although it is unlikely that you will find Cal Sets with
these names.
4.
Verify that the analyzer is in Multiport mode. See the bottom of the measurement
window.
a.
If only four S-Parameters are listed, select
Utility > System > Configure > Multiport Capability
-OR-
(Instrument > Setup External Hardware > Multiport > Multiport Configuration...)
.
Select
Restart as a multiport PNA with this test set
and select
U3022AM06
(8-Port)
from the drop-down menu > OK. Refer to
5.
Press
[Preset]
.
6.
Verify that the
[Start Frequency]
is set to
[10 MHz]
.
7.
Set the
[Stop Frequency]
is set to
[26.5 GHz]
or
[Upper Freq]
limit for your analyzer.
8.
Select
[Power] > Power Level
and enter
[0 dBm]
.
9.
Select
Response > Avg > IF Bandwidth > 100 Hz > OK
.
If you are using an E8362B,
select
[Sweep Setup] > Bandwidth > [100 Hz] >[Enter]
.
10.
Select
Stimulus > Sweep > Number of Points > 401
.
If you are
using
an E8362B,
select
[Sweep Setup] > Points > [401] > [Enter]
.
11.
Connect the ECal module to an available USB port on the front or rear panel of the
analyzer. This procedure assumes you are using a ECal. If you are not, see
Calibration and Verification Procedure” on page
12.
Allow the ECal module, Test Set
,
and analyzer to warm up for a minimum of 30
minutes.
72
Keysight U3022AM06 User's and Service Guide