U3022AM06
General Specifications
General Specifications
When connected to an Analyzer, this Test Set will degrade the performance at the
test ports. The internal switch paths reduce test port power and power to the
receivers. This affects the test port power of the Analyzer and also reduces dynamic
range. The reflection tracking values measured in the
can be subtracted from the analyzers dynamic range to determine the
approximate performance of the system.
Power Requirements
Verify that the required ac power is available before installing the Test Set to the
Analyzer.
•
100/120/220/240 VAC (50/60Hz)
•
The instruments can operate with mains supply voltage fluctuations up to ± 10%
of the nominal voltage.
•
Air conditioning equipment (or other motor–operated equipment) should not be
placed on the same ac line that powers the Test Set to the Analyzer.
•
U3022AM06 maximum power is 350 W.
This is a Safety Class I product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted into a socket
outlet provided with a protective earth contact. Any interruption of the protective
conductor, inside or outside the instrument, is likely to make the instrument
dangerous. Intentional interruption is prohibited.
Specifications for the Test Set are characteristic for the system performance of the
Analyzer and Test Set. Actual performance of the system is based on the customer’s
Analyzer and options that are used with the Test Set. A functional certificate is only
offered for the Test Set.
An N-Port calibration should be performed for optimum measurement accuracy.
A periodic calibration is not required. The Operators Check should be performed after
System Setup, or if performance is in question.
The life expectancy of the following switches are five million cycles. When making
measurements using more than one port, single sweeps are recommended to
minimize switch cycles.
• 87104-60045 (switch, x4)
• N1811-60027 (switch x1, Option 129 only)
Keysight U3022AM06 User's and Service Guide