![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 71](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380071.webp)
U3022AM06
Cal Kit Operational Check
Verification Limits
Typical performance is based on 1 to 2 units, refer to
Table 20
. System performance
for the analyzer and Test Set are only characteristic and intended as non-
warrantied information. Only a functional certificate is provided for the Test Set.
It is recommended that you return your instrument to Keysight Technologies for
servicing or repair if the Test Set and analyzer performance exceed the operational
verification limits.
A periodic calibration is not required. The Operators Check should be performed after
System Setup, or if performance is in question. An N-Port calibration should be
performed before making a measurement.
Table 20 Reflection Tracking Limits
Frequency
E8362B/C &
N5230A/C
N5242A
/B
10 MHz to 4 GHz
-5 dB
-5 dB
4 GHz to 6 GHz
-6 dB
-6 dB
6 GHz to 10 GHz
-8 dB
-8 dB
10 GHz to 15 GHz
-12 dB
-12 dB
15 GHz to 20 GHz
-14 dB
-14 dB
20 GHz to 26.5 GHz
-15 dB
a
a. Reflection tracking takes into account Source Loss, Receiver Loss, Margin, and analyzer Mixer Cal.
Keysight U3022AM06 User's and Service Guide
71