7 Run Power Calibration
To retrieve and set the VISA Address of the Power Meter (Power Sensor)
1. Open Keysight Connection Expert.
2. Select the Instruments tab.
a. On the left side of the dialog box, from the list of USB instruments,
select the U8487A Power Sensor or equivalent.
b. On the right side of the dialog box, under "Connection Strings",
highlight the VISA Address of the Power Sensor
and select Ctrl-C (to copy the VISA Address to the Clipboard); this should
be of the form:
USB0::0x2A8D::0xA618::MY59160002::0::INSTR
c. Open the S910xA System Calibration dialog,
select the Equipment tab,
click the VISA Address entry box,
and type Ctrl-V (to paste the VISA Address from the Clipboard).
d. Once the VISA Address has been entered, verify connectivity
by clicking the Test button located next to the entry box;
Address Valid
is displayed on a button if the address connects properly.