7 Run Power Calibration
S9100A 5G Multi-Band Vector Transceiver, Startup Guide
71
To retrieve and set the VISA Address of the Primary or Secondary Transceiver
1. Open Keysight Connection Expert.
2. Select the Instruments tab.
a. On the left side of the dialog box, from the list of LAN instruments,
select the M9410A, Keysight Technologies
localhost
.
b. On the right side of the dialog box, under "Connection Strings",
highlight the VISA Address to be used on the S910xA System Calibration
dialog Equipment tab.
c. On the S910xA System Calibration dialog, select the Equipment tab.
d. Once the VISA Address has been entered, verify connectivity
by clicking Test located next to the VISA Address entry box;
Address Valid
is displayed on a button if the address connects properly.