Keysight N5290/1A Installation Guide 4-15
Performance Verification
System Verification
shows an example of typical verification results with Print Graphs
selected in the Printer Output area of the System Verification dialog box. The
printed graphical results show the following:
— the name of the device measured
— the serial number of the device
— the parameters measured
— Results of the measurements.
— Data measured at the factory from the verification kit.
— Upper and lower limit points as defined by the total system uncertainty
system.
Figure 4-6
Example of Printed Graphical Verification Results (4-port is Shown)