4-10 Keysight N5290/1A Installation Guide
Performance Verification
System Verification
4-
Performing System Verification
The following verification procedure is automated by the analyzer firmware.
The process for the verification is:
— connect cables to the analyzer test ports
— perform a calibration or recall a recent calibration
— run the system verification program for the verification devices
Each time through the verification process, you are prompted to make
necessary connections and perform or recall a calibration as part of performing
the verification. If you select to perform a calibration, you are guided through
the calibration procedure. This part of the process can be eliminated if you
choose to load an existing recent calibration that was created by the
verification process. If necessary, refer to the analyzer’s on-line embedded help
for information on storing and recalling calibrations.
For each verification device, the analyzer reads a file from the verification
media and sequentially measures the magnitude and phase for all four
S-parameters.
Equipment Used in the System Verification Procedure
Cable Substitution
The test port cables specified for the network analyzer system have been
characterized for connector repeatability, magnitude and phase stability with
flexing, return loss, insertion loss, and aging rate. Since test port cable
For system verification to perform correctly, it is NECESSARY that the
verification devices be measured with their female connectors connected
to port 1 or 3 and their male connectors connected to port 2 or 4.
Although the performance for all S-parameters are measured, the S
11
and
S
22
phase uncertainties for the attenuators and airlines are less important
for verifying system performance. Therefore, the limit lines will not appear
on the printout.
Equipment
Type
1.0 mm
Calibration kit
85059B
Verification kit
85059V
Cables
Included in the
85059B Kit
Adapters
None required