3- 12 Keysight N5290/1A Installation Guide
System Installation
PNA, Controller, and Test Head Module Interconnections
3-
10.
Install Interconnect Kit Semirigid Cables:
Figure 3-7
To avoid damaging your RF ports 1/2/3/4, always use two wrenches
IMPORTANT!
The N5227B and N5247B interconnect kits contain 2.4 mm to 3.5 mm
cables. It is important to connect the 2.4 mm end to the PNA and to
connect the 3.5 mm end to the N5292A test set controller. Improper
connections will result in damage to your instruments.
It is important to use two wrenches when connecting the semirigid cables
to the 1/2/3/4 port 1.85 mm outputs front ports of the PNA to avoid
stressing the connectors and causing damage to your PNA (i.e., the
2.4 mm cable end is compatible with the 1.85 mm connector of the PNA).