Keysight N5290/1A Installation Guide 3-19
System Installation
PNA, Controller, and Test Head Module Interconnections
N5293/5A Modules Front Panel Cabling
The front-panel connections between the millimeter head controller and a test
head module are shown in
through
. This cabling is duplicated for each test head module in the system.
—
“N5292A Option 200 2–Port Cable Connections” on page 3-20
—
“N5292A Option 400 4–Port Cable Connections” on page 3-21
Connections for Each Head
1.
Connect a set of x4 feet to the bottom of each millimeter module.
Choose either:
—
N5293-40003 feet, bottom, grip or
—
N5293-40004 feet, bottom, slip
2. PNA/PNA-X
: Verify you have connected all of the semirigid cables and
torque to 8 in-lbs. Refer to
.
3. N5292A
: Connect the integrated cables from the N9293A or N5295A
millimeter heads.
IMPORTANT!
In order to maintain compliance to the certificate of calibration it is
recommended that the frequency extenders be connected to the
corresponding test set ports per the Certificate of Calibration provided.
IMPORTANT!
Before connecting the millimeter-wave modules, verify that the test set
and the power supplies (if used) are powered down.
The N5292A test set controller’s integrated front panel connectors contain
optical sensors to ramp power up or down gradually when the N5293/5A
millimeter modules are connected to prevent damage. Refer to
“Troubleshooting the N5293/5A Millimeter Modules” on page 3-26
IMPORTANT!
When connecting the N5293/5A modules, it is very
important to verify that the module connectors are seated flush and that
the screws are screwed all the way into the N5292A front test ports. But,
be careful to
not
overtighten.
IMPORTANT!
In order to maintain compliance to the certificate of calibration it is
recommended that the frequency extenders be connected to the
corresponding test set ports per the Certificate of Calibration provided.