3- 6 Keysight N5290/1A Installation Guide
System Installation
PNA, Controller, and Test Head Module Interconnections
3-
PNA, Controller, and Test Head Module Interconnections
System Setup
Mount the analyzer on top of the test set before connecting any cables to the
PNA or to the test set. The cables can then be connected after the analyzer
and test set are securely attached.
If not already done, open the interconnect kit and validate its contents against
.
Mounting a Network Analyzer on the Test Set
Refer to the following kits for your specific analyzer. In this procedure a various
PNAs are used with a N5292A Test Set Controller. But, other PNAs are similar.
— Interconnect kits includes the SMA 50 ohm loads, cables, upper and lower
locking feet, and screws. Refer to
.
IMPORTANT!
The N5227B and N5247B interconnect kits contain 2.4 mm to 3.5 mm
cables. It is important to connect the 2.4 mm end to the PNA and to
connect the 3.5 mm end to the N5292A test set controller. Improper
connections will result in damage to your instruments.
It is important to use two wrenches when connecting the semirigid cables
to the 1/2/3/4 port 1.85 mm outputs front ports of the PNA to avoid
stressing the connectors and causing damage to your PNA (i.e., the
2.4 mm cable end is compatible with the 1.85 mm connector of the PNA).
The network analyzer is heavy. It is recommended that two
individuals, or a mechanical lift be used to lift or transport the
instrument.