Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
56
The M9037A embedded controller is recommended for multi-chassis
configurations. The M9036A embedded controller or standalone PC are both
supported. Not all standalone personal computers work with four chassis
configuration. For a list of personal computers that have been tested to be
compatible with the M9018/M9019A chassis, refer to
. Similarly, using M9037A or M9036A
with PXI/AXIe Chassis – Technical Overview
has several prerequisites. For more information, please contact your local Keysight
Technologies Inc. representative.
The Keysight M9021A PCIe cable interface modules are connected from one
chassis to the other, as shown in the image below. In the M9381A chassis group,
the Sync connector of the first M9311A PXIe Digital Vector Modulator of each
chassis must be connected to each other. Similarly, in the M9391A chassis group,
the Sync connector of the first M9214A PXIe Digitizer of each chassis must be
connected to each other. Multi-channel operation requires proper configuration of
the chassis backplane trigger lines. For more information, refer to
M9018A PXIe Backplane Triggers Configuring M9018B/M9019A PXIe Backplane
. The PXIe VSGs or PXIe VSAs can also be configured to be phase coherent
with Option 012 and by cabling all channels in the chassis group to use a single
M9301A PXIe Synthesizer module via the
. See the
V2802A LO Distribution Network
image below. Slot blockers and EMC filler panels are installed in the empty slots.
The recommended 8x8 system is configured as follows:
Chassis #1
VSG #1:
M9381A
VSG #2:
M9381A
REF:
M9300A
VSG #3:
M9381A
VSG #4:
M9381A
M9021A PCIe Cable
Interface in slot 1
M9300A in
slot 10