The Keysight M9018A PXIe Chassis does not remember the trigger bus
configuration upon cycling power. After a power cycle, you must reconfigure
the bus triggers by using the M9018A SFP or by creating a persistent
configuration by using the AgM9018 API. Refer to the Keysight M9391A PXIe
VSA and M9381A PXIe VSG Programming Guide (M9300-90080) for details and
programming examples. This document is available on your installation CD and
online at
http://edadownload.software.keysight.com/Modular/M938XA/2.0
.
/M9391_and_M9381_ProgrammingGuide.pdf
Procedure for Configuring PXIe Backplane Bus Triggers
Make sure you have followed Steps 1 through 4 in this Startup Guide to
setup and configure the Keysight M9381A PXIe Vector Signal Generator.
Conduct a Self Test (
). If your
Utilities > Self Test... > Run Self Test
configuration straddles two slot groups, you will receive error messages.
In the following graphic a Self Test has been run on an M9381A that is
installed in slots 5 through 9; therefore, it straddles two slot groups. The
Self Test detects errors with backplane triggers 6 and 7.
Open the M9018A SFP (
Start > All Programs > Keysight > M9018 > M9018
).
SFP
Select (enable)
.
Allow Control