Holding the module by the injector/ejector handle, slide it into an available
PXIe (or hybrid) slot, as shown in the following figure.
Install the module into the slot of the chassis by placing the module
card edges into the front module guides (top and bottom).
Slide the module to the rear of the chassis and ensure that the injector
/ejector handle is pushed down in the unlatched (downward) position.
Slide the module completely into the chassis. When you begin to feel
resistance, pull up on the injector/ejector handle to fully inject the
module into the chassis.
Secure all modules to the chassis using the module mounting screws. Use a
#1 Pozidriv® or a slotted screwdriver and torque to 5 Lb-In (0.57 N-m).
Performance may suffer if the screws are not tightened properly.
Verify that the PXIe chassis fans are operable and free of dust and other
contaminants that may restrict airflow.
Install filler panels and slot blockers after installing the modules. Missing
filler panels or slot blockers may disrupt air circulation in the chassis. The
left-most slot does not accept a slot blocker.
Your shipment included an Keysight M9310A PXIe Source Output and an
Keysight M9350A PXIe Downconverter. With these modules you received a
straight, coaxial adapter (1250-2316) (SMA female to SMA male). Connect
this adapter to the RF Out connector of each M9310A and the RF In
connector of each M9350A, and torque to 8 Lb-In (0.904 Nm).
This will
.
extend the life of these often-used connectors
Use the
M9391A Instrument Connections with M9381A VSG
cables to the instrument. The torque specification for SMA connectors is 8
Lb-In (0.904 Nm).
If you are using a PCIe Cable Interface, such as the Keysight M9021, connect
the Cable Interface in the chassis to the PC host per the instructions that
came with the Cable Interface.
Power up the PXIe chassis.