Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
5.
See
for positioning the M9391A modules.
M9391A Instrument Connections
See
for positioning the M9381A modules.
M9381A Instrument Connections
The M9381A uses peer-to-peer (module-to-module) triggering across
the PXIe Express chassis backplane. Therefore, be sure to install the
M9381A modules into a chassis that supports peer-to-peer PXIe
Express I/O switch topology. When using the Keysight M9018/M9019B
PXIe 18-Slot Chassis, install the M9381A modules into one of its three
slot groups: 1 through 6, 7 through 12, or 13 through 18. Unless you
install the M9381A modules entirely within one of the slot groups, you
will experience trigger errors. If you install the modules across
(straddling) any of these groups of slots, you will have to reconfigure the
chassis's PXIe trigger bus routing for trigger 6 and trigger 7.
(EXCEPTION: You can install the M9300A Reference module in a
different slot group, thus allowing one M9300A module to provide an RF
reference across slot groups to one or more modular instruments.) For
details on reconfiguring the M9018/M9019A chassis triggers, see
Configuring M9018A PXIe Backplane Triggers Configuring M9018B
/M9019A PXIe Backplane Triggers
documentation Keysight M9018B Chassis documentation Keysight
. If you are using a non-Keysight
chassis, consult your manufacturer's documentation.
When using the Keysight M9018/M9019A Chassis you will achieve the
highest data throughput for the Keysight M9391A PXIe Vector Signal
Analyzer by using Slots 2, 6, 11 or 15 for the Keysight M9214A PXIe IF
Digitizer. They are "x8"slots and the others are."x4". For further details
see the
/
Keysight M9018A Chassis documentation Keysight M9018B
. If you
Chassis documentation Keysight M9019A Chassis documentation
are using a non-Keysight chassis, consult your manufacturer's
documentation.