31
Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
14.
15.
16.
Remap the backplane triggers, if necessary. See
Backplane Triggers Configuring M9018B/M9019A PXIe Backplane Triggers
If you are using a remote PC, reboot the PC host.
Proceed to
Step 5 - Verify Operation of the M9391A VSA and M9381A VSG
M9381A Instrument Connections
Single M9381A VSG
This section contains a cabling diagram for the Keysight M9381A PXIe Vector
Signal Generator, and a cable and module association table. For front panel feature
descriptions, see the following:
If you are building an M9381A PXIe VSA plus an M9391A PXIe VSG, see
.
M9391A and M9381A Instrument Connections
The recommended single PXIe VSG system is configured as follows:
VSG: M9381A
REF: M9300A
M9311A in slot 2/3
M9310A in slot 4
M9301A in slot 5
M9300A in slot 10
Install the semi-rigid cables first.
Recommendation:
Torque specification
for all SMA connectors is 8 Lb-In (0.904 Nm).
M9381A Cable and Module Table