Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
28
1.
2.
3.
4.
Step 4 - Install the Module
Step 4 - Install the Modules
PXIe hardware does not support "hot-swap" (changing modules while power is
applied to the chassis) capabilities. Before installing or removing a module to
/from the chassis, power off the chassis to prevent damage to the module.
All component modules for an M9381A or M9391A instrument are factory
tested, aligned, calibrated and shipped as a "bundle". It is important that you
maintain the bundle when installing modules or returning an instrument for
repair.
These modules can be used in a chassis with PXIe or PXIe-H chassis slots.
Module Installation Process Overview
The module can be installed in any PXIe or hybrid PXIe slot marked with a peripheral slot
compatibility image (solid black circle for PXIe, or solid black circle with the letter -H- for hybrid).
Make sure that the line cord is plugged in to establish earth ground and that
the chassis power switch is
.
Off
If the chassis has multiple fan speed settings, ensure that the fan switch is
set to
and the inhibit switch is set to
.
AUTO
DEF
Position the chassis to provide ample space between the chassis fan intake
and exhaust vents. Blockage by walls or obstructions affects the air flow
needed for cooling. (Refer to the chassis documentation for more information
about cooling).
Before inserting a module into the chassis, back the mounting screws out to
ensure that there is no interference between the screws and the mounting
rails.