![Keysight Technologies M9381A PXIe Startup Manual Download Page 45](http://html1.mh-extra.com/html/keysight-technologies/m9381a-pxie/m9381a-pxie_startup-manual_1962005045.webp)
45
Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
The Keysight M9021A PCIe cable interface is extended from one chassis to the
other. The Sync connector of the first M9214A PXIe Digitizer of the master chassis
must be cabled to the Sync connector of the first M9214A PXIe Digitizer of the
other chassis. The PXIe VSAs in each chassis can be time-synchronized to measure
multi-channel transmitters and analyze MIMO devices. Multi-channel operation
requires proper configuration of the chassis backplane trigger lines. For more
information, refer to
Configuring M9018A PXIe Backplane Triggers Configuring
. The PXIe VSAs can also be configured
M9018B/M9019A PXIe Backplane Triggers
to be phase coherent with Option 012 and by cabling all channels to use a single
M9301A PXIe Synthesizer module via the
. Slot
V2802A LO Distribution Network
blockers and EMC filler panels are installed in the empty slots.
The recommended multiple PXIe VSA system is configured as follows:
Chassis 1:
VSA #1: M9391A
VSA #2: M9391A
REF: M9300A
VSA #3: M9391A
VSA #4: M9391A
M9214A in slot 2
M9301A in slot 3
M9350A in slot 4
M9214A in slot 6
M9301A in slot 7
M9350A in slot 8
M9300A in slot 10
M9214A in slot 11
M9301A in slot 12
M9350A in slot 13
M9214A in slot 15
M9301A in slot 16
M9350A in slot 17
Chassis 2: