37
Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
VSG #1: M9381A
VSG #2: M9381A
REF: M9300A
VSG #3: M9381A
VSG #4: M9381A
M9311A in slot 2
/3
M9310A in slot 4
M9301A in slot 5
M9311A in slot 6
/7
M9310A in slot 8
M9301A in slot 9
M9300A in slot
10
M9311A in slot 11
/12
M9310A in slot 13
M9301A in slot 14
M9311A in slot 15
/16
M9310A in slot 17
M9301A in slot 18
Chassis 2:
VSG #5: M9381A
VSG #6: M9381A
REF: M9300A
VSG #7: M9381A
VSG #8: M9381A
M9311A in slot 2
/3
M9310A in slot 4
M9301A in slot 5
M9311A in slot 6
/7
M9310A in slot 8
M9301A in slot 9
M9300A in slot
10
M9311A in slot 11
/12
M9310A in slot 13
M9301A in slot 14
M9311A in slot 15
/16
M9310A in slot 17
M9301A in slot 18
Other module and instrument placements can be used, but this configuration
optimizes speed and makes cabling simpler for phase coherent configuration.
The M9300A frequency Reference module must be installed in slot 10 for
proper time synchronization between instruments. For more information, see
.
Sharing the M9300A Frequency Reference
Cabling is different for time-synchronization-only (independent LO) configuration
and phase-coherent (shared LO) configuration.
Time-Synchronized Only, Multi-Channel PXIe VSG (Independent LO)
When phase coherence is not required, cable the instrument as follows for time
synchronized multi-channel operation for independently tunable channels.
8x8 Independent LO Cable Connections:
– Primary 100 MHz connection (8121-2175, long SMB-SMB); chassis
sync; 10 MHz distribution
– Secondary 100 MHz connections (8120-5091, short SMB-SMB);
trigger cables
– LO distribution connections; Semi-rigids (W1312-20266 and W1312-
20267)
– PCIe cable (Y1202A)