Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
92
3.
4.
5.
6.
Module
Green
Orange
Red
Off
determined
until the
power
supply
failure is
resolved.
M9350A
The Soft Front Panel
has initialized the
connection to the
module
Missing 100 MHz
Reference.
n/a
Not
connected
by the SFP.
Failure in
the power
supplies.
Module
hardware
health
can't be
determined
until the
power
supply
failure is
resolved.
It is important that no signal is present at the RF Input of the Keysight
M9350A PXIe Downconverter when doing a Self Test. If a signal is
present, it may result in a false failure.
If the M9381A is configured for Shared LO, run the LO Level Alignment on
the M9381A designated as master, then run the alignment on the M9381A's
designated as slaves (
).
Utilities > LO Level Alignment > All States
Run the alignments on each M9391A (
Utilities > Alignments >
). If the instrument is configured for Shared LO,
Comprehensive > All States
you must run the alignments on the M9391A designated as master before
you run the alignments on the M9391A's designated as slaves.
Conduct a Self Test on all modules of the M9391A and also the M9381A (
).
Utilities > Self Test... > Run Self Test
If the Self Test passes (see results below), go to
.