Step 6 - Make a Measurement
As a best practice, always run a Self Test of each instrument before making a
measurement.
If your solution has more than one channel, repeat this process for each
channel. Alternatively, you can use Signal Studio to make measurements with
multiple M9381A instruments and the 89600 VSA software to make
measurements with multiple M9391A instruments (see
).
The first step in this process is to open the SFP of the M9381A and the
M9391A.
Open the M9381A SFP by selecting
Start > All Programs > Keysight >
.
M938x > M9381 SFP
Open the M9391A SFP by selecting
Start > All Programs > Keysight >
.
M9391 > M9391 SFP
For each SFP, you are presented with the "Connect to Instrument"
dialog. Use
to select all of the modules that are
Ctrl/Select
components of the M9381A and the and press
.
Connect
Connect a high quality SMA (male) to SMA (male) cable between the RF Out
connector on the Keysight M9310A PXIe Source Output and the RF In
connector on the Keysight M9350A PXIe Downconverter.
Torque the connectors to 8 Lb-In (0.904 Nm).
On the M9381A SFP make the following settings:
Frequency: 2 GHz
Amplitude: -10 dBm
RF On: checked
ALC On: checked.