Keysight M9391A PXIe Signal Analyzer & M9381A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
90
2.
c.
3.
If not all modules and slot locations are visible in the SFP
"Connect to Instrument" dialog, Close the SFP. See
Selecting
initializes the instrument and all associated
Connect
modules. Prior to initialization, all LEDs are off.
Check the front panel Status LEDs of all modules, according to the following
table.
Module
Green
Orange
Red
Off
M9300A
The Soft Front Panel
has initialized the
connection to the
module
n/a
Indicates that the
VCXO is unlocked.
Not
connected
by the SFP.
Failure in
the power
supplies.
Module
hardware
health
can't be
determined
until the
power
supply
failure is
resolved.
M9301A
The Soft Front Panel
has initialized the
connection to the
module
Tuning is in progress,
or the M9301A is
unlocked from the
reference.
n/a
Not
connected
by the SFP.
Failure in
the power
supplies.
Module
hardware
health
can't be
determined
until the
power
supply
failure is
resolved.
M9310A
n/a
n/a