3-20
Keysight B2900 User’s Guide, Edition 5
Installation
Connecting a DUT
CAUTION
Maximum voltage and current
The test fixture and adapters must be used under the following limitations to prevent
damage.
Keysight N1294A-001:
250 V maximum,
42 V maximum for connecting
N1295A
Keysight N1294A-002:
250 V maximum,
42 V maximum for connecting
N1295A
Keysight N1295A:
42 V, 1.05 A maximum
Using the 16442B Test Fixture
Keysight 16442B is a test fixture which has six triaxial connectors for
source/measure unit (SMU), the GNDU connector for the ground unit of a
parameter/device analyzer, the Intlk connector for the interlock control, and six
coaxial connectors for other instruments. See
Keysight 16442B User’s Guide
for
details and accessories of the 16442B.
Requirements
•
Keysight 16442B test fixture furnished with pin clip wire, pin plug wire, etc.
•
Keysight N1294A-001 banana to triaxial adapter for 2-wire connections
•
Keysight N1294A-002 banana to triaxial adapter for 4-wire connections
•
Triaxial cable, 2 ea. for 2-wire connections, and 3 ea. for 4-wire connections
NOTE
Shielding
16442B is equipped with a cover. To minimize the effect of ambient noise, close the
cover when performing measurements.
CAUTION
Maximum voltage and current
The test fixture and adapters must be used under the following limitations to prevent
damage.
Keysight N1294A-001:
250 V maximum,
200 V maximum for connecting
16442B
Keysight N1294A-002:
250 V maximum,
200 V maximum for connecting
16442B
Keysight 16442B:
200 V, 1 A maximum for SMU input
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......