4-28
Keysight B2900 User’s Guide, Edition 5
Front Panel Reference
Config key group
Output Connection dialog box
This dialog box provides the following parameters for setting the channel operation
and connection.
Ch
Only on 2-channel models. Channel 1 (Ch 1) or
2 (Ch 2)
This field specifies the channel set by this
dialog box.
Sensing Type
Sensing type, 2-wire connection (2-WIRE) or
4-wire connection (4-WIRE). Set 4-WIRE for
remote sensing.
Low Terminal State
Low sense terminal connection,
grounded (GROUND) or floating (FLOAT)
High Capacitance Mode
High capacitance mode, ON or OFF
Set the mode to ON to perform high capacitive
load measurements. See
Over Voltage/Current Protection
Over voltage or current protection, ON or OFF
Set the function ON to turn off the channel
output automatically and immediately when it
reaches its compliance.
Output-Off State
Output-off state, high impedance (HIGH Z),
normal (NORMAL), or zero volt (ZERO)
This is the source setup condition after output
off. See
Auto Output-On
Automatic output on function, ON or OFF
Set the function ON to turn on the channel
output automatically just before the trigger
system is initiated by the SCPI command (not
by front panel operation).
Auto Output-Off
Automatic output off function, ON or OFF
Set the function ON to turn off the channel
output automatically and immediately when all
trigger system changes the status from busy to
idle.
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......