Keysight B2900 User’s Guide, Edition 5
3-7
Installation
Installing the Keysight B2900
Storage
10 % to 90 %RH, non-condensing
•
Altitude
Operating
0 m to 2,000 m (6,561 ft.)
Storage
0 m to 4,600 m (15,092 ft.)
•
Power requirement: 100-240 V (±10 %), 50/60 Hz
•
Maximum volt-amps (VA): 250 VA
•
Dimensions
Without handle and feet: 88 mm (H)
213 mm (W)
450 mm (D)
Bench configuration (with handle and feet): 180 mm (H)
260 mm (W)
480
mm (D)
Fan cools the instrument by drawing air through the sides and exhausting it out the
back. The instrument must be installed in a location that allows sufficient space at
the sides and back of the instrument for adequate air circulation.
Connecting the Power Cord
WARNING
FIRE HAZARD: Use only the power cord supplied with your instrument.
Using other types of power cord may cause overheating of the power cord,
resulting in fire.
SHOCK HAZARD: The power cord provides the chassis ground through a
third conductor. Be sure to connect to a three-conductor type power outlet with
the correct pin grounded.
RISQUE D’INCENDIE : utilisez uniquement le câble d'alimentation fourni
avec votre appareil. L’utilisation d’autres types de câble d’alimentation peut
provoquer une surchauffe du câble d’alimentation et provoquer un incendie.
RISQUE DE CHOC ÉLECTRIQUE: le câble d’alimentation fournit la masse
du châssis par le biais d’un troisième conducteur. Assurez-vous de connecter la
prise d'alimentation de type trois conducteurs avec la broche correcte mise à la
terre.
NOTE
The detachable power cord may be used as an emergency disconnecting device.
Removing the power cord will disconnect AC input power to the instrument.
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......