Keysight B2900 User’s Guide, Edition 5
6-13
Function Details
Output-Off Status
Output-Off Status
The output-off status is the condition which is set on the source/measure unit (SMU)
automatically, immediately after its output is turned off. The status must be specified
before enabling the source output. The available conditions are shown in
To set the output-off status, see
“Output Connection dialog box” on page 4-28
Table 6-1
Output-Off State
NOTE
This setting is not applied to the output-off process triggered by the emergency
condition such as the over voltage/current protection, interlock open, and over
temperature protection. Then the output voltage is immediately set to 0 V and the
output switch is set to off.
Name
Conditions after output-off
HIGH Z,
high
impedance
•
Output relay: off (open or break)
•
Voltage source setup is not changed if the source applies
40 V or less.
•
Current source setup is not changed if the source uses the
100 mA range or lower.
NORMAL
•
Source function: Voltage source
•
Output voltage: 0 V
•
Current compliance: 100
A at the 100
A range
•
Output relay: off (open or break)
ZERO
•
Source function: Voltage source
•
Output voltage: 0 V
•
Current compliance: 100
A at the 100
A range
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......