3-24
Keysight B2900 User’s Guide, Edition 5
Installation
Installing the Interlock Circuit
Installing the Interlock Circuit
The interlock circuit is a simple electric circuit, as shown in
. The circuit
electrically opens when an access door is opened, and closes when the door is
closed.
B2900 cannot apply high voltages over
42 V when the Digital I/O interlock
terminal is open. To perform high voltage measurements, the B2900 interlock
terminal must be connected to the interlock circuit installed in the measurement
environment such as the shielding box. The interlock circuit is important and
necessary to prevent electrical shock when the user touches the measurement
terminals.
WARNING
Hazardous voltage, instrument maximum output voltage may appear at the
High Force, High Sense, and Guard terminals if the fixture cover is closed.
To prevent electrical shock, do
not
expose these lines.
Une tension dangereuse, une tension de sortie maximale de l'appareil peut
apparaître aux bornes High Force, High Sense et Guard si le couvercle de
l'équipement est fermé. Afin d’éviter toute décharge électrique, n’exposez pas
ces lignes.
Figure 3-11
Interlock Circuit
Requirements
•
LED (Keysight part number 1450-0641 or equivalent), 1 ea.
•
Mechanical switch (Keysight N1254A-402 or equivalent), 2 ea.
•
Resistor, 500
to 1 k
(choose lower one to get higher output of LED), 1 ea.
Mechanical switches
Shielding box
LED
Access door
D-sub connector
Interlock circuit
B2900
Digital I/O
15, 19
14
17, 18
16
24, 25
Dsub cable
15, 19
22, 23
16, 17
24
14
Negative
Positive
or
22
R
R: 500
W
to 1 k
W
25
R
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......