High Voltage Shock Hazard
Keysight B2900 can force dangerous voltages (±210 V) at the High Force, Guard, and
High Sense terminals. To prevent electric shock hazard, the following safety precautions
must be observed during the use of Keysight B2900.
• Use a three-conductor AC power cable to appliance coupler (inlet) and the
instrument to an electric ground (safety ground).
• Prepare shielding box which covers interface to a device under test and equipped
with interlock circuit that opens when the door is opened.
• Before performing measurement, connect the interlock circuit to the Interlock
terminal of this instrument.
• Confirm periodically that the interlock function works normally.
• Before touching the connections of the High Force, Guard, and High Sense terminals,
turn the instrument off and discharge any capacitors of the measurement path. If you
do not turn the instrument off, complete “all” of the following items, regardless of any
instrument’s settings.
• Terminate source output by pressing the On/Off switch, confirm that the On/Off
switch turns off.
• Confirm that the HV (high voltage) status indicator is not lit.
• Open the shielding box access door (open the Interlock terminal).
• Discharge any capacitors if the capacitance is connected to this instrument.
• Warn workers in the vicinity of the instrument about hazardous conditions.
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......