2-14
Keysight B2900 User’s Guide, Edition 5
Introduction
Source/Measure Unit
Table 2-4
Limitations for using Channels 1 and 2
Table 2-5
Current Measurement Range
Channel 1 voltage V1
Channel 2 voltage V2
Current limit
a
a. I1: Channel 1 current, I2: Channel 2 current
0 < |V1|
6 V
0 < |V2|
6 V
I1 + I2
4 A
6 V < |V2|
21 V
I1 + I2
1.6
4 A
6 V < |V1|
21 V
0 < |V2|
6 V
I1 + I2
0.625
2.5 A
6 V < |V2|
21 V
I1 + I2
2.5 A
Range value
Current measurement value
Resolution
10 nA
a
a. Available for the B2911A and B2912A. Not available for the B2901A
and B2902A.
0
|I|
10.6 nA
10 fA
100 nA
0
|I|
106 nA
100 fA
1
A
0
|I|
1.06
A
1 pA
10
A
0
|I|
10.6
A
10 pA
100
A
0
|I|
106
A
100 pA
1 mA
0
|I|
1.06 mA
1 nA
10 mA
0
|I|
10.6 mA
10 nA
100 mA
0
|I|
106 mA
100 nA
1 A
0
|I|
1.06 A
1
A
1.5 A
0
|I|
1.53 A
3 A
0
|I|
3.06 A
10
A
10 A
b
b. Available for pulse mode. Not available for DC mode.
0
|I|
10.6 A
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......