To Execute Limit Test
To Display Limit Test Result
For details on the setup parameters, see
“Limit Test Result dialog box” on page
1. Press the Result > Limit Test function keys.
The Limit Test Result dialog box opens.
2. In the Ch field, set the channel for the data to be displayed.
The limit test result is displayed in the area below the Length field.
3. Press OK to close the dialog box.
The Length field displays the number of data.
In the area below the Length field, the limit test result is displayed in the following
format:
(aaaaa) BIN: bb DATA: +c.ddd
(aaaaa)
Data index
aaaaa
BIN:
Bin number
bb
(
01
to
12
)
The bin numbers 1 to 12 are associated with the index numbers
1 to 12. See
“To Set Individual Limit Tests” on page 5-29
If the limit test data is out of the bins,
00
is set for the
GRADING mode and
15
is set for the SORTING mode.
DATA:
Limit test data
+c.ddd