BOP-1K 080709
4-13
2. Press
!
- VOLTAGE,
!
- INTERNAL,
@
- ZERO to set the BOP to zero volts output.
Connect a Digital Voltmeter (DVM) to the BOP OUT S and COM S terminals to measure the
output voltage. Adjust the BOP output as needed until the DVM reads as close to zero as
possible within tolerance specified in Table 4-3 for VOLTAGE ZERO. Press
%
once.
3. Press
@
- POSITIVE. Measure the voltage output using the DVM. Adjust as needed until
the DVM reading is as close as possible above the nominal full scale value within the toler-
ance specified in Table 4-3 for +FULL SCALE VOLTAGE. Press
%
once.
4. Press
#
- NEGATIVE to set the BOP to maximum negative output voltage and adjust as
needed until the reading is as close as possible above the nominal full scale value within the
limits specified in Table 4-3 for –FULL SCALE VOLTAGE. Press
%
twice.
5. Press
$
- PROTECT,
@
- POSITIVE to adjust the maximum positive voltage protection
limit of the power supply while working in current mode. Adjust as needed until the reading is
as close as possible above the nominal full scale value within the limits specified in Table 4-3
for +FULL SCALE VPR LIMIT. Press
%
once.
6. Press
#
- NEGATIVE to adjust the maximum negative protection limit of the power supply.
Adjust as needed until the reading is as close as possible above the nominal full scale value
within the limits specified in Table 4-3 for –FULL SCALE VPR LIMIT. Press
%
three times.
7. Press
#
- EXTRL CONTROLS,
$
- EXTERN PROTECT,
!
- ZERO to set the BOP to
zero volts output. Connect 0V ±0.1mV d-c reference voltage from the external voltage source
to the following pins of analog I/O connector (A2A5J6) referenced to pin 12 (GND1):
VLIM+_EXTERNAL (pin 14), VLIM–_EXTERNAL (pin 6), ILIM+_EXTERNAL (pin 13), ILIM–
_EXTERNAL (pin 5). Then press
!
- ZERO.
This
enables the BOP to recognize that the
voltage at the four _EXTERNAL pins (14, 6, 13, and 5) is 0V. Press
%
once.
NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to
meet published specifications.
8. Replace 0V ±0.1mV d-c reference voltage at pins 14, 6, 13, and 5 with +10.0V ±0.1mV d-c
reference voltage from the external voltage source referenced to pin 12 (GND1) of the ana-
log I/O connector (A2A5J6), then press
@
- MAXIMUM. This enables the BOP to recognize
that the voltage at the four _EXTERNAL pins (14, 6, 13, and 5) is 10.0V. Then press
%
three times.
9. Disc10.0V d-c reference voltage from pins 14, 6, 13 and 6 of analog I/O connector
(A2A5J6). Connect a +0V ±0.1mV d-c reference voltage to EXT_REF, pin 11, referenced to
pin 12 (GND1), of the BOP analog I/O port (A2A5J6).
10.Press
!
- VOLTAGE,
#
- EXTERNAL,
!
- ZERO to set the BOP to zero volts output.
Adjust as needed
until the DVM reads as close to zero as possible within the limits specified
in Table 4-4 for VOLTAGE ZERO. Press
%
twice.
11. Press
@
- GAIN,
!
- ZERO to set the BOP to zero volts output. Adjust as needed
until the
DVM reads as close to zero as possible within the limits specified in Table 4-4 for VOLTAGE
ZERO. Press
%
twice.
12.C10.0V ±0.1mV d-c reference voltage from the external voltage source to
EXT_REF, pin 11, referenced to pin 4 (SGND) of Analog I/O Port connector (A2A5J6), Press
@
- GAIN,
@
- POSITIVE. Adjust as needed
until the DVM reads as close to zero as pos-
sible within the limits specified in Table 4-3 for VOLTAGE ZERO. Press
%
twice.
Summary of Contents for BOP 1000W
Page 2: ......
Page 10: ......
Page 20: ......
Page 31: ......
Page 32: ......
Page 36: ......
Page 50: ...FIGURE 2 2 LOAD CONNECTIONS LOCAL SENSING...
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ......
Page 57: ......
Page 64: ......
Page 68: ...FIGURE 3 3 POWER UP SCREEN SHOWING GRAPHIC METERS VOLTAGE 0000 0...
Page 125: ......
Page 128: ......