B-28
BOP-1K 080709
B.105 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:NEGative? QUERY
VOLT:PROT:LIM:NEG?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:NEG?
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:NEGative?
Returns: <exp_value>
Description:Identifies the minimum (maximum absolute) value possible for the negative protection limit
for voltage.
B.106 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive COMMAND
VOLT:PROT:LIM:POS
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:POS <exp_value>
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive <exp_value>
where <value> is between minimum (box) value (Figure 1-3) and 1% above rated (nominal) output
voltage.
Description:Establishes the maximum value possible for the positive protection limit for voltage. This
may be set up to 1% greater than rated voltage. Requires MEM:UPD (PAR. B.12) to save for power
up. See PAR. 3.5.2.2 and Figure 3-16 for special programming considerations.
B.107 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive? QUERY
VOLT:PROT:LIM:POS?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:POS?
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive?
Returns: <exp_value>
Description:Identifies the maximum value possible for the positive protection limit for voltage.
B.108 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude] COMMAND
VOLT:TRIG
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP] <exp_value>
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1
Description:
Programs voltage value to be transferred to output by *TRG command.
Actual output voltage will
depend on load conditions. If the value exceeds the maximum for the model being programmed, error
message -222,”Data out of range” is posted in output queue. If value exceeds VOLT:LIM:HIGH value,
a value corresponding to the voltage limit will be programmed.
B.109 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]? QUERY
VOLT:TRIG?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]?
Return Value:
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1
Description:
Returns value representing voltage value to be programmed by *TRG command established by
VOLT:TRIG command).
B.110 STATus:OPERation:CONDition? QUERY
STAT:OPER:COND?
Syntax:
Short Form: STAT:OPER:COND?
Long Form: STATus:OPERation:CONDition?
Return Value:
<int_value>
Description:
Returns the value of the Operation Condition Register (see Table B-3).
The Operation Condition
Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-10.)
TABLE B-3. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,
AND OPERATION EVENT REGISTER BITS
CONDITION
NU
LR
NU
LC
SC
CC
TC
CV
NU
TA WTG
NU
BIT
15
14
13
12
11
10
9
8
7
6
5
4 - 0
VALUE
32,768 16,384 8192 4096 2048 1024 512 256 128
64
32
16 -1
CC
Constant Current Mode
CV Constant
Voltage
Mode
LC
List Complete
LR
List Running
NU
Not Used
SC
Sample Complete
TC
Transient Complete
TA
Transient Armed
WTG Waiting
For
Trigger
Summary of Contents for BOP 1000W
Page 2: ......
Page 10: ......
Page 20: ......
Page 31: ......
Page 32: ......
Page 36: ......
Page 50: ...FIGURE 2 2 LOAD CONNECTIONS LOCAL SENSING...
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ......
Page 57: ......
Page 64: ......
Page 68: ...FIGURE 3 3 POWER UP SCREEN SHOWING GRAPHIC METERS VOLTAGE 0000 0...
Page 125: ......
Page 128: ......