3-4
BOP HIPWR 080709
FIGURE 3-3. POWER-UP SCREEN SHOWING GRAPHIC METERS
FIGURE 3-4. POWER-UP SCREEN SHOWING TIME LINE GRAPH
When in Voltage mode, the power supply will (within the configured and rated limits) provide the
programmed output voltage. Current is determined by the load, and cannot exceed the Current
Protect limits. Mode displayed at the upper right is VOLTAGE SOURCE. If the protect limit is
reached, CPROTECT is displayed.
When in Current mode, the power supply will (within the configured and rated limits) provide the
programmed output current. Voltage is determined by the load, and cannot exceed the Voltage
Protect limits. Mode displayed at the upper right is CURRENT SOURCE. If the protect limit is
reached, VPROTECT is displayed.
If CPROTECT (current protection), VPROTECT (voltage protection) or PROTECT (analog pro-
tection, either voltage or current) is displayed, the power supply output is controlled by the pro-
tect channel and will not exceed the preprogrammed protection limit.
When in Source mode, the power supply is delivering (sourcing) energy to a passive load.
VOLTAGE
0000
0.
VOLTAGE
0000
0.