Section 7: Command reference
Models 707B and 708B Switching Matrix Reference Manual
7-150
707B-901-01 Rev. A / August 2010
scan.mode
This attribute controls the scan mode setting.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
System reset
Scan reset
Create
configuration script
scan.MODE_OPEN_ALL
Usage
scanModeSetting = scan.mode
scan.mode = scanModeSetting
scanModeSetting
The present scan mode setting. Set to one of the following values:
•
scan.MODE_OPEN_ALL
or
0
: Opens all relays on all slots before a scan
starts
•
scan.MODE_OPEN_SELECTIVE
or
1
: Opens relays intelligently; see
Details
Details
When this attribute is set to
scan.MODE_OPEN_ALL
, all channels on all slots are opened before a scan starts.
When this attribute is set to
scan.MODE_OPEN_SELECTIVE
, an intelligent open is performed:
•
All channels involved in scanning are opened
•
Closed channels not involved in scanning remain closed during the scan
Example
scan.mode = scan.MODE_OPEN_SELECTIVE
Sets the scan mode setting to open
selective.
Also see
(on page 7-150)
scan.reset()
This function resets the trigger model and scan list settings to their factory default settings.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
scan.reset()
Details
When
scan.reset()
is sent, the trigger model and scan settings that are reset to the factory defaults are:
•
scan.bypass
•
scan.mode