Section 2: General operation
Models 707B and 708B Switching Matrix Reference Manual
2-82
707B-901-01 Rev. A / August 2010
Instrument control
To reset the entire TSP-enabled system, including the controlling node and all subordinate nodes,
click
Reset
.
Create a script using TSB Embedded
NOTE
If you are using TSB Embedded to create scripts, you do not need to use the commands
loadscript
or
loadandrunscript
and
endscript
. For information on using TSB Embedded,
select the Help button on a web page or the Help option from the navigational pane on the left hand
side.
You can create a script in from the instrument web page with TSB Embedded. When you save the
script in TSB Embedded, it is loaded into the runtime environment and saved in the nonvolatile
memory of the instrument.
To create a script using TSB Embedded:
1. In the TSP Script box, enter the name of the script.
2. In the input area, enter the sequence of commands to be included in the script. Line numbers are
automatically assigned.
3. Click
Save Script
. The name is added to the User Scripts list on the left.
Admin page
Through the Admin page, you can change the instrument password and the instrument time.
Change the password
To change the password for the web interface:
1. In the web interface, from the left navigation, click
Admin
. A login page is displayed.
2. Enter the current password in the Password box. (The default is "admin".)
3. Click
Login
. The Administration page is displayed.
4. In the Current Password box, enter the current password.
5. In the New Password box, enter the new password.
6. In the Confirm New Password box, enter the new password again.
7. Click
Submit
. The new password takes effect immediately.