4-16
Sweeps
Model 6220/6221 Reference Manual
Return to
Running a log staircase sweep
1.
Configure source functions.
Examples
– The following commands restore defaults, set the bias current
to 10
μ
A, and the compliance to 5V:
2.
Configure the sweep.
Examples
– The following commands configure a single logarithmic
staircase sweep from 1mA to 10mA with five points using a 2s delay, best
fixed source range, and compliance abort disabled:
3.
Arm and run the sweep:
4.
When the sweep is done, turn the source output off with this
command:
OUTP OFF
*RST
‘ Restore 622x defaults.
SOUR:CURR 1e-5
‘ Set bias current to 10
μ
A
SOUR:CURR:COMP 5
‘ Set compliance to 5V.
SOUR:SWE:SPAC LOG
‘ Select log staircase sweep.
SOUR:CURR:STAR 1e-3
‘ Set start current to 1mA.
SOUR:CURR:STOP 1e-2
‘ Set stop current to 10mA.
SOUR:SWE:POIN 5
‘ # points = 5.
SOUR:DEL 2
‘ Set delay to 2s.
SOUR:SWE:RANG BEST
‘ Select best fixed source range.
SOUR:SWE:COUN 1
‘ Set sweep count to 1.
SOUR:SWE:CAB OFF
‘ Disable compliance abort.
SOUR:SWE:ARM
‘ Arm sweep, turn on output.
INIT
‘ Trigger sweep.