14-18 SCPI Reference Tables
Model 6220/6221 Reference Manual
Return to
Section 14 topics
:ILINe?
Query TRIG input signal line.
:OLINe <NRf>
Set TRIG output signal line: 1 to 6.
2
:OLINe?
Query TRIG output signal line.
:OUTPut <name>
Enable/disable TRIG output trigger:
SOURce[1], DELay, or NONE.
NONE
:OUTPut?
Query TRIG output trigger.
Table 14-11
Units command summary
Command
Description
Default
Sec
SCPI
UNIT
Commands to set reading units for Delta, Pulse Delta,
and Differential Conductance:
[:VOLT]
Volts readings are sent from 2182/2182A to 622x.
[:DC] <name>
Specify reading units*: V, OHMS, W, or SIEMens.
V
[:DC]?
Query reading units.
:POWer
Power reading type for Pulse Delta (6221):
[:TYPE] <name>
Set power units reading type: PEAK or AVERage.
AVER
[:TYPE]?
Query power units reading type.
* The <name> parameter for Siemens can be sent as S, SIEM or SIEMENS.
Table 14-10 (cont.)
Trigger command summary
Command*
Description
Default
Sec
SCPI
* Arm layer commands control entire sweep. Trigger Layer commands control individual sweep steps.