5-48
Delta, Pulse Delta, and Differential Conductance
Model 6220/6221 Reference Manual
Return to
The previous read command reads the last Pulse Delta reading that was
performed by the Model 2182A. If this command is sent before a new
reading is available, the last Pulse Delta reading will again be returned.
7.
When finished with Pulse Delta, it can be disarmed by sending the following
command:
8.
Recall stored Pulse Delta readings
– Model 2182A Pulse Delta readings
were sent to the buffer of the Model 6221. Send the following read com-
mand to read the buffer:
NOTE
If a read command is sent when Pulse Delta is
not running, error -221 Settings Conflict will oc-
cur. See
for details on errors associat-
ed with Pulse Delta operation.
NOTE
Details on using Model 6221 “
”
are provided on
.
SOUR:SWE:ABOR
‘
‘
Stops Pulse Delta and places the Model 2182A
in the local mode.
TRACe:DATA?
‘ Read Pulse Delta readings stored in 6221 buffer.