17-10 Calibration
Model 6220/6221 Reference Manual
Return to
9.
Press ENTER. The display will prompt:
-20V COMPLIANCE CAL
Press ENTER to Output -2.0000mA
10.
Press ENTER to output -2mA for -20V compliance calibration. The instru-
ment will prompt as follows:
DMM RDG=-020.0000V
Use , , , ,
ENTER,
or
EXIT
11.
Note the DMM reading, then adjust the Model 6220/6221 display to agree
with that value.
12.
Press ENTER. The display will prompt:
-.1V COMPLIANCE CAL
Press ENTER to Output -2.0000mA
13.
Press ENTER to output -2mA for -0.1V compliance calibration. The instru-
ment will prompt as follows:
DMM RDG=-0.100000V
Use , , , ,
ENTER,
or
EXIT
14.
Note the DMM reading, then adjust the Model 6220/6221 display to agree
with that value.
15.
Press ENTER. The display will prompt:
100V COMPLIANCE CAL
Press ENTER to 2.0000mA
16.
Press ENTER to 2mA for 100V compliance calibration. The instru-
ment will prompt as follows:
DMM RDG=+100.0000V
Use , , , ,
ENTER,
or
EXIT
17.
Note the DMM reading, then adjust the Model 6220/6221 display to agree
with that value.
18.
Press ENTER. The display will prompt:
21V COMPLIANCE CAL
Press ENTER to 2.0000mA
19.
Press ENTER to 2mA for 21V compliance calibration. The instru-
ment will prompt as follows:
DMM RDG=+21.00000V
Use , , , ,
ENTER,
or
EXIT
20.
Note the DMM reading, then adjust the Model 6220/6221 display to agree
with that value.
21.
Press ENTER. The display will prompt:
-100V COMPLIANCE CAL
Press ENTER to Output -2.0000mA
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...