Model 6220/6221 Reference Manual
Output Connections
2-17
Return to
Figure 2-12
Noise shield
Figure 2-13
Safety shield
Cable Guard connections
Triax Cable Guard is used to provide guarding for the triax cable and can be
extended all the way to the DUT at a metal guard plate. Connections for Cable
Guard are shown in
.
A safety shield must be used whenever hazardous voltages (
>
30Vrms, 42Vpeak)
will be present in the test circuit.
shows a how to make guarded con-
nections with the use of a safety shield.
High
Low
DUT
Output
Metal noise shield
High
Low
DUT
Output
Metal noise shield
622x
622x
(red*)
(black*)
(red*)
(green*)
* Boot color for alligator clip if
using supplied triax cable.
A) Inner shield connected to Output Low
B) Inner shield connected to Cable Guard,
Output Low connected to Earth Ground
(Earth Ground)
Earth Ground
High
Low
DUT
Output
Safety
Earth
Ground
Metal
safety
shield
High
Low
DUT
Output
Safety
Earth
Ground
Metal
safety
shield
Chassis
Screw
622x
622x
(red*)
(black*)
(red*)
(green*)
(green*)
* Boot color for alligator clip if
using supplied triax cable.
A) Inner shield connected to Output Low
B) Inner shield connected to Cable Guard,
Output Low connected to earth ground
(Earth Ground)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com